Agilent Technologies, Inc. 72 SATA PHY, TSG & OOB Test MOI v.1.1 Revision 1.4
C.2 – Measurement Results
Figure C.2a shows the measured single-ended return loss of the laboratory load, showing that the response
meets the specified limits of >20dB return loss from 100MHz to 5 GHz, and >10 dB from 5GHz to 8GHz for both
channels 1 and 3 on the DSO81304B. Figure C.2b shows the measured differential return loss of the laboratory load.
Figure C.2a Single-ended return loss measurement for Agilent’s DSO81304B laboratory load (including
Rosenberger SMA cables and 11742A DC blocking capacitors)
Figure C.2a Differential return loss measurement for Agilent’s DSA91204A laboratory load (including Rosenberger
SMA cables and 11742A DC blocking capacitors)